posted on 2022-12-16, 09:50 authored by James E Sharman, Alberto P. Avolio, Johannes Baulmann, Athanase Benetos, Jacques Blacher, C Leigh Blizzard, Pierre Boutouyrie, Chen-Huan Chen, Philip Chowienczyk, John R. Cockcroft, J. Kennedy Cruickshank, Isabel Ferreira, Lorenzo Ghiadoni, Piotr Jankowski, Stephane Laurent, Barry McDonnell, Carmel McEniery, Sandrine C. Millasseau, Theodoros G. Papaioannou, Gianfranco Parati, Jeong Bae Park, Athanase D Protogerou, Mary J. Roman, Giuseppe Schillaci, Patrick Segers, George S. Stergiou, Hirofumi Tomiyama, Raymond R. Townsend, Luc M. Van Bortel, Jiguang Wang, Siegfried Wassertheurer, Thomas Weber, Ian B. Wilkinson, Charalambos Vlachopoulos

The original Riva-Rocci method to measure blood pressure (BP) using a cuff at the upper arm assumed the pressure obtained by this technique was a good proxy for central aortic BP.1,2 The clinical (prognostic) importance of brachial cuff BP is undeniable for both the assessment of cardiovascular risk associated with elevated BP and the benefits of treatment-induced BP reduction.3 However, it is also generally appreciated that peripheral artery systolic BP (SBP; brachial or radial artery) may be an inaccurate substitute for central SBP.4 This has been reported in human studies using intra-arterial catheterization of peripheral and central arteries.5–8 There may also be a discrepancy between peripheral and central BP responses to vasoactive drugs.9 These findings are corroborated in larger studies using non-invasive central aortic BP methods,10–13 and, while yet to be fully adopted in clinical practice, an independent prognostic value of central BP has been demonstrated.14–16 Altogether, there is a growing interest among clinicians towards improving risk estimates by using devices that provide more accurate measures of central aortic BP than those provided by current brachial cuff BP methods.

Many non-invasive devices have been developed that purport to estimate central BP from different peripheral artery sites (e.g. radial, brachial, carotid arteries) using different principles of recording the pressure or surrogate signals (e.g. applanation tonometry, oscillometry, ultrasound, or magnetic resonance imaging) and different calibration methods to derive central BP. Since upper arm cuff-based devices to estimate central BP are more clinically appealing, in recent years several companies have developed such devices using a variety of techniques (e.g. oscillometric sub-diastolic or supra-systolic waveform analysis with generalized transfer functions), which employ a variety of signal processing steps to estimate central BP from peripheral signals.17,18 Yet, with no standardized guidelines,17 the accuracy testing of these new devices (as well as the preceding devices) has not been undertaken in a uniform fashion with comparable protocols, emphasizing the need for guidance in this field.19–22 An international task force was convened to address this situation.

History

Published in

European Heart Journal

Publisher

Oxford University Press

Version

  • VoR (Version of Record)

Citation

Sharman, J.E., Avolio, A.P., Baulmann, J., Benetos, A., Blacher, J., Blizzard, C.L., Boutouyrie, P., Chen, C.H., Chowienczyk, P., Cockcroft, J.R. and Cruickshank, J.K., et al. (2017) 'Validation of non-invasive central blood pressure devices: ARTERY Society task force consensus statement on protocol standardization', European Heart Journal, p.ehw632.

Print ISSN

0195-668X

Electronic ISSN

1522-9645

Cardiff Met Affiliation

  • Cardiff School of Sport and Health Sciences

Cardiff Met Authors

John R. Cockcroft Barry J. McDonnell

Cardiff Met Research Centre/Group

  • Cardiovascular Physiology

Copyright Holder

  • © The Authors

Language

  • en

Usage metrics

    Cardiovascular Health & Ageing - Journal Articles

    Categories

    Licence

    Exports

    RefWorks
    BibTeX
    Ref. manager
    Endnote
    DataCite
    NLM
    DC